Multiphoton laser scanning microscope
Multiphoton or 2-photon microscopy (MPM/2PM) combines confocal resolution with in depth imaging. Typically, it is used to image multicellular systems like the roots / leaves of plants, or other strongly scattering media. As special features, it is equipped with spectral detectors, resonant scanners, fluorescence lifetime imaging (FLIM) detection and image deconvolution.
When femtosecond pulsed near infrared (NIR) light is focused to a diffraction limited spot of ~200 nm diameter, there is a probability that the energy of the photons is combined to excite a molecule. For example, two NIR photons of 800 nm can be simultaneously absorbed for a transition that would normally require the energy of a single 400 nm photon.
Technical details
For detection, the layout is different from CLSM. As multiphoton excitation is only possible in the focal spot of the objective, there is no need to place a pinhole before the detector (and descan the optical beam). Therefore, a dichroic mirror is placed just below the objective to reflect the emission directly to one of two Hybrid detectors. The optical configuration and the size of the detectors is such that the loss of photons due to scattering (i.e. by changes in refractive index) is minimized.
Multiphoton microscopy is combined with fluorescence lifetime imaging (FLIM). FLIM monitors the distribution of the fluorescence lifetimes of a fluorophore at the different locations within the sample. The fluorescence of a sample is monitored as a function of time after excitation by a flash of light. The lifetime may be sensitive to environmental factors like ion concentration, pH and polarity but is independent of dye concentration or light path length. The lifetime can be significantly reduced if excited state processes like FRET occur. Moreover, it can be used to measure the ultrafast dynamics of photosynthesis in leaves, algae or bacteria.
Applications
The combination with FLIM allows for Förster Resonance Energy Transfer (FRET) imaging, as well as pH, viscosity, and polarity imaging. Note that the use of an MPM requires more training than a regular CLSM. High energy pulsed laser light (CLASS 4) is used that is almost invisible but can cause severe damage to the human eye. The detectors are very sensitive and less well protected than in a regular confocal. You can contact Arjen Bader or Jan Willem Borst for an introduction.
In addition to the Multiphoton laser scanning microscope, Wageningen University & Research, Sharing Research Facilities offers the use of: